測汞儀:
F732-V, F732-VJ, F732-S, F732-G, ZYG-11, DHL-2, DHL-3, DHL-4, DHL-100, DHL-200, DHL-300, HL-2D, HL-6D, HL-7D, HL-2B, HL-3B, HL-4B, HL-5B, HL-6B, HL-7B,HL-100B, HL-200B, HL-300B, HL-600B, HL-2, HL-3, HL-4, HL-5, HL-6, HL-7, HL-100, HL-200, HL-300, HL-600, DHL-A, HDL-B, HL-1, HL-1B, HL-1S, HL-2, HL-2B, HL-2D, HL-2S, HL-3, HL-3B, HL-4, HL-4B, HL-4D, HL-5, HL-5A, HL-5B, HL-6, HL-6B, HL-6D, BT-100, BT-100B, BT-100D, BT-200, BT-200B, BT-600, BT-600B, T-600C, HL-2N, BT-100N,
測汞儀(yí)(mercury vapourmeter,mercury vapour analyzer)
是(shì)一(yī)種高靈敏度(dù)的測汞用的原子(zǐ)吸(xī)收光譜(pǔ)的儀器。在一些金(jīn)屬礦床上方空氣中的汞異常往往(wǎng)低(dī)到(dào)幾(jǐ)至幾十納克/立方(fāng)米。原有各種(zhǒng)測汞的方(fāng)法無法發現此種微弱(ruò)異常。近年來研究成功的測汞儀,其靈(líng)敏度可以(yǐ)達到l納(nà)克/立方米。它(tā)是利用(yòng)汞蒸(zhēng)氣能強烈(liè)吸收253.7納米譜(pǔ)線的特性而設計的(de)。儀器主要包(bāo)括發射(shè)253.7納米譜線的汞燈,氣體吸(xī)收室(shì)及光(guāng)電(diàn)放(fàng)大(dà)和測量(liàng)等裝置。進入吸收室的氣體(tǐ)樣(yàng)品,如含(hán)有微跡的(de)汞,則(zé)通過吸收室的光線會因部分被汞吸收而減(jiǎn)弱。根(gēn)據光線減(jiǎn)弱的程(chéng)度(dù)可以測出氣體(tǐ)中的汞含量。二氧化硫及許多稀有氣體253.7納米譜線鄰近(jìn)。對(duì)譜(pǔ)線都顯著的吸收。因而產生嚴重的幹擾。由於消除幹擾方法的(de)不同而設計製成了多種類型的測汞(gǒng)儀。例如(rú):①利用貴金屬捕集器使汞被截留,使幹擾氣體(tǐ)逸去;②使樣品氣流分(fèn)成兩股(gǔ),將一股中的(de)汞事先(xiān)移(yí)除,然後比較同(tóng)一光源通過兩個吸(xī)收室時的輸出;③利(lì)用壓致展(zhǎn)寬效應,將通過吸收室後的光線分成兩股,一(yī)股(gǔ)再通過(guò)飽和汞蒸氣(qì)室,然後測(cè)量兩股(gǔ)透出光線強(qiáng)度的(de)比值;④利用“塞曼(màn)效應(yīng)”,比較在光源上施加(jiā)磁(cí)場與不加磁場時,通過吸(xī)收室的光線強度的比值。根據實用上的要求,已製(zhì)成了裝在汽車及飛機上進行連續測(cè)定汞的儀器,以及(jí)能就(jiù)地進(jìn)行測定的輕便(biàn)背包式的測(cè)汞儀器等。
技(jì)術參數:
1.檢測下限:≤0.01ng/ml
2.線性相關係數(shù):r>0.999
3.相對標準偏差:<4%
4.製(zhì)造商:大吉光電
5.測量(liàng)範圍:0-10.0ug/L
6.穩定度(dù):讀數漂(piāo)移≤±2字/3分鍾(在A=0處)
7.精密度(dù):變異係數≤4% |