測汞儀:
F732-V, F732-VJ, F732-S, F732-G, ZYG-11, DHL-2, DHL-3, DHL-4, DHL-100, DHL-200, DHL-300, HL-2D, HL-6D, HL-7D, HL-2B, HL-3B, HL-4B, HL-5B, HL-6B, HL-7B,HL-100B, HL-200B, HL-300B, HL-600B, HL-2, HL-3, HL-4, HL-5, HL-6, HL-7, HL-100, HL-200, HL-300, HL-600, DHL-A, HDL-B, HL-1, HL-1B, HL-1S, HL-2, HL-2B, HL-2D, HL-2S, HL-3, HL-3B, HL-4, HL-4B, HL-4D, HL-5, HL-5A, HL-5B, HL-6, HL-6B, HL-6D, BT-100, BT-100B, BT-100D, BT-200, BT-200B, BT-600, BT-600B, T-600C, HL-2N, BT-100N,
測汞儀(mercury vapourmeter,mercury vapour analyzer)
是一種高靈敏度的測汞用的原子吸收光譜的儀器。在一(yī)些金屬礦床上方(fāng)空氣中(zhōng)的汞異常(cháng)往往低到幾(jǐ)至幾十納克/立方(fāng)米。原有(yǒu)各種(zhǒng)測汞的方法無法發現此種(zhǒng)微弱異(yì)常。近年來研(yán)究(jiū)成功(gōng)的測(cè)汞儀,其靈敏度可(kě)以(yǐ)達到(dào)l納克/立方(fāng)米。它是利用汞蒸氣能強烈吸收253.7納米譜線的特性(xìng)而設計的。儀器主(zhǔ)要(yào)包(bāo)括(kuò)發(fā)射253.7納米譜(pǔ)線的汞燈,氣體吸(xī)收室及光電放大和測量等裝置。進入吸收室的氣體(tǐ)樣品(pǐn),如含有(yǒu)微跡的汞,則通過(guò)吸(xī)收室的光線會因部分被汞吸收而減弱。根據光線(xiàn)減弱的程度可以測出氣體中的汞含量。二氧化(huà)硫及許多稀有氣體253.7納米譜線(xiàn)鄰近(jìn)。對譜線都顯著的吸收。因而產(chǎn)生嚴(yán)重的幹擾。由於消除幹擾方法的不同而設(shè)計製(zhì)成了多種類型的測(cè)汞儀。例如:①利用貴金屬捕集器使(shǐ)汞被截留,使幹(gàn)擾氣體逸去;②使樣品(pǐn)氣流分成兩股,將一股中的(de)汞事先移除,然後比較同一光源通過兩個吸收室時的(de)輸出(chū);③利(lì)用壓致展寬效應,將通(tōng)過吸收室後的光線分(fèn)成兩股,一股(gǔ)再通過飽和汞蒸氣(qì)室,然後測量兩股透出光線強度的比值(zhí);④利用“塞曼效應”,比(bǐ)較在光源上施(shī)加磁場與不加(jiā)磁場(chǎng)時,通過吸收(shōu)室的光線(xiàn)強度的比值。根(gēn)據實用(yòng)上的要求,已製成了裝在汽車及飛機上進行連續測定汞的儀器,以及能就地(dì)進行測(cè)定的輕便背包式(shì)的測汞儀器等。
技術(shù)參數:
1.檢測下限:≤0.01ng/ml
2.線性相關(guān)係數:r>0.999
3.相對標準偏(piān)差:<4%
4.製造商(shāng):大吉光電
5.測量範(fàn)圍:0-10.0ug/L
6.穩定(dìng)度:讀數漂移(yí)≤±2字/3分鍾(在A=0處)
7.精密度:變異係數≤4% |